Electronic Components Datasheet Search
  English  ▼

Delete All
ON OFF
ALLDATASHEET.COM

X  

Preview PDF Download HTML

M9340A Datasheet(PDF) 10 Page - Keysight Technologies

Part No. M9340A
Description  M9485A PXIe Multiport Vector Network Analyzer
Download  40 Pages
Scroll/Zoom Zoom In 100% Zoom Out
Maker  KEYSIGHT [Keysight Technologies]
Homepage  http://www.keysight.com
Logo 

M9340A Datasheet(HTML) 10 Page - Keysight Technologies

Zoom Inzoom in Zoom Outzoom out
 10 / 40 page
background image
10 | Keysight | M9485A PXIe Multiport Vector Network Analyzer 1 MHz/50 MHz to 9 GHz - Data Sheet
M9485A PXIe Multiport Vector Network Analyzer Characteristics (Typical)
Option 2xx (Combination of M9377A and M9378A/B Configuration)
Corrected system performance
Description
Typical
System dynamic range at test port3
(Source: Standard Configuration, Receiver: Standard Configuration, Test Attenuator 0 dB IF Bandwidth = 10 Hz)
10 M to 50 MHz
65 dB
50 M to 100 MHz
122 dB
100 M to 200 MHz
128 dB
200 M to 400 MHz
134 dB
400 M to 800 MHz
140 dB
800 M to 3 GHz
144 dB
3 G to 5 GHz
141 dB
5 G to 6 GHz
138 dB
6 G to 7 GHz
135 dB
7 G to 9 GHz
134 dB
(Source: Standard Configuration, Receiver: Wide Dynamic Range Configuration, Test Attenuator 0 dB IF Bandwidth = 10 Hz,)
10 M to 50 MHz
118 dB
50 M to 3 GHz
160 dB
3 G to 5 GHz
157 dB
5 G to 6 GHz
154 dB
6 G to 7 GHz
150 dB
7 G to 9 GHz
149 dB
Extended Dynamic Range at direct receiver access input 2, 3
(Source: Standard Configuration, Receiver: Direct Access Configuration, Tin, Test Attenuator 0 dB, IF Bandwidth = 10 Hz)
10 M to 50 MHz
119 dB
50 M to 3 GHz
162 dB
3 G to 5 GHz
159 dB
5 G to 6 GHz
156 dB
6 G to 7 GHz
153 dB
7 G to 9 GHz
152 dB
1. Thetestportsystemdynamicrangeiscalculatedasthedifferencebetweenthetestportrmsnoisefloorandthesourcemaximumoutputpower.
Crosstalk does not limit the dynamic range.
2. Theextendeddynamicrangeiscalculatedasthedifferencebetweenthedirectaccessreceiverinputrmsnoisefloorandthesourcemaximumoutput
power. Crosstalk does not limit the dynamic range. The extended dynamic range is only available in one-path transmission measurements.
3. The effective dynamic range must take measurement uncertainties and interfering signals into account.


Html Pages

1  2  3  4  5  6  7  8  9  10  11  12  13  14  15  16  17  18  19  20  21  22  23  24  25  26  27  28  29  30  31  32  33  34  35  36  37  38  39  40 


Datasheet Download




Link URL




Privacy Policy
ALLDATASHEET.COM
Does ALLDATASHEET help your business so far?  [ DONATE ]  

About Alldatasheet   |   Advertisement   |   Datasheet Upload   |   Contact us   |   Privacy Policy   |   Alldatasheet API   |   Link Exchange   |   Manufacturer List